Product Description
4 Gas H2S CO CH4 O2 Calibration Gas – 116L Cylinder
This four-gas calibration mixture is used to confirm detector performance where oxygen deficiency and combustible hazards must be verified together. The reduced oxygen concentration activates alarm response while toxic and LEL channels are validated in a single exposure.
The 116L cylinder supports daily testing routes without frequent cylinder replacement.
Gas Concentrations
| Component | Concentration |
|---|---|
| Hydrogen Sulfide (H₂S) | 50 PPM |
| Carbon Monoxide (CO) | 50 PPM |
| Methane (CH₄) | 2.5% (50% LEL) |
| Oxygen (O₂) | 12.0% |
| Balance | Nitrogen |
Cylinder Specifications
| Property | Value |
|---|---|
| Cylinder Size | 116L (4.0 ft³) |
| Cylinder Specification | C-10 |
| Cylinder Material | Aluminum |
| Gas Reactivity | Reactive |
What This Cylinder Verifies
-
Toxic gas detection response (H₂S & CO)
-
Oxygen-deficiency alarm activation
-
Combustible gas response (methane LEL)
Used for confirming alarm thresholds before entering hazardous environments.
Typical Applications
-
Confined space entry programs
-
Daily bump testing routes
-
Industrial compliance inspections
-
Wastewater monitoring
-
Preventative maintenance calibration
Custom Mixtures Available
Alternate concentrations and alarm setpoints available upon request.