H₂S 25 PPM / CO 100 PPM / CH₄ 2.5% (50% LEL) / O₂ 20.9% 4-Gas Calibration Gas – 34L Disposable Cylinder
This 4-gas calibration blend contains:
-
Hydrogen Sulfide (H₂S) – 25 PPM
-
Carbon Monoxide (CO) – 100 PPM
-
Methane (CH₄) – 2.5% (50% LEL)
-
Oxygen (O₂) – 20.9%
-
Balance Gas – Nitrogen
Designed for portable multi-gas monitor calibration and bump testing under standard atmospheric oxygen conditions.
Gas Composition
H₂S 25 PPM / CO 100 PPM / CH₄ 2.5% (50% LEL) / O₂ 20.9%
Balance: Nitrogen
Gas Count: 4
Cylinder Specifications
Cylinder Size: 34L (1.2 Cubic Feet)
Cylinder Pressure: 500 PSI
Valve Type: C-10
Cylinder Material: Aluminum
Gas Reactivity: Reactive
Cylinder Type: Disposable
The 34L 500 PSI configuration is optimized for:
-
Mobile safety technicians
-
Smaller confined space programs
-
Lightweight field kits
-
Lower-volume calibration cycles
Application Positioning
This blend supports:
-
Routine OSHA-required bump testing
-
Standard atmospheric verification
-
Detector fleet spot-checking
-
Utility and municipal maintenance
Oxygen at 20.9% simulates ambient air conditions.
Methane at 2.5% equals 50% LEL for combustible sensor validation.
CO at 100 PPM aligns with common alarm setpoint testing requirements.
Why Choose the 34L 500 PSI Format?
-
Compact and easy to transport
-
Lower upfront cost than larger cylinders
-
Ideal for limited-use or rotating compliance schedules
-
Compatible with standard C-10 fixed-flow regulators